ISO81060-2 :2013,(Non-Invasive Sphygmomanometers - Part 2: Clinical Validation Of
Automated Measurement Type).
ELECTROMAGNETIC COMPATIBILITY INFORMATION
This product is applicable to the equipment and system requirements for the purpose of
receiving radio frequency energy for the purpose of the work, Bluetooth receive bandwidth
2M. This product can also be used to include RF transmitter equipment and system
requirements and emission frequency of 2.4GHz ISM band, Bluetooth modulation
types:GFSK, effective max radiated power: <4dBm
Phenomenon
RF emissions
Harmonic distortion
Voltage
fluctuations
and flicker
Phenomenon
Electrostatic
Discharge
Radiated RF EM field
Proximity fields from
RF
wireless
communications
equipment
Rated
power
frequency
magnetic
fields
Table 3 – Proximity fields from RF wireless communications
Test frequency
Band
(MHz)
(MHz)
385
380-390
450
430-470
710
704-787
745
780
Table 1 - Emission
Compliance
CISPR 11
Group 1, Class B
IEC 61000-3-2
Class A
IEC 61000-3-3
Compliance
Table 2 - Enclosure Port
Basic EMC standard
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-3
IEC 61000-4-8
equipment
Pulse modulation 18Hz, 27V/m
FM, ±5kHz deviation, 1kHz sine, 28V/m
Pulse modulation 217Hz, 9V/m
13 / 14
Electromagnetic environment
Home healthcare environment
Home healthcare environment
Home healthcare environment
Immunity test levels
Home Healthcare Environment
±8 kV contact
±2kV, ±4kV, ±8kV, ±15kV air
10V/m
80MHz-2.7GHz
80% AM at 1kHz
Refer to table 3
30A/m
50Hz or 60Hz
Immunity test levels
Professional healthcare facility environment