13
Activate the
checkboxes
corresponding
to the active
channels on the
multiplexer.
When a particular channel runs, the amber LED for that
channel illuminates.
NOTE: If you are running AC experiments (e.g., EIS),
you must run AC calibration of your potentiostat with
the multiplexer connected. For all DC experiments
calibration is not needed.