Olympus DP23 사용 설명서 - 페이지 23
{카테고리_이름} Olympus DP23에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Olympus DP23 31 페이지. Microscope digital camera
Problem: A snapshot cannot be acquired
Possible cause
DP2-TWAIN, cellSens or
OLYMPUS Stream is still
processing or performing data
storage after acquisition.
The computer's RAM is fully
occupied.
There are too many images
loaded in your software.
Problem: The image is too bright
Possible cause
The exposure compensation is
set too high to the + side.
A dark area of the image was
mistakenly set as the metering
area
The AE lock is set at an exposure
time that is too long.
The illumination of the
microscope is too bright.
The manual exposure time is too
long.
Remedy
Wait until the processing is finished. Then
acquire the next snapshot.
Close other software applications. Then
acquire the next snapshot.
If necessary, save the images. Close the
loaded images. Then acquire the next
snapshot.
Remedy
Set the exposure compensation to a
more suitable value.
Set a more suitable metering area.
Deactivate the AE lock.
Reduce the illumination light level or use
an attenuation filter.
Reduce the manual exposure time.
7 Troubleshooting
23