Olympus BH-2 BHT 매뉴얼 - 페이지 12

{카테고리_이름} Olympus BH-2 BHT에 대한 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Olympus BH-2 BHT 24 페이지. Sliding focus block
Olympus BH-2 BHT에 대해서도 마찬가지입니다: 완전한 분해, 청소 및 재조립 (23 페이지)

Olympus BH-2 BHT 매뉴얼
 
AC Power Switch 
The AC power switch (see Figure 11) is an Alps SDT‐7 chassis‐mount DPST rocker switch with a TV‐5 rating for UL/CSA.  
This  switch  cannot  be  cleaned  with  control/contact  cleaner  since  there  are  no  openings  to  inject  the  cleaner.        The 
mounting holes of this switch are spaced 34mm on center. 
 
Figure 11 – AC Power Switch Mounted to BHT/BHTU Electrical Base (100/115V Version) 
Voltage Selector Switch 
The voltage selector switch (see Figure 12) is a chassis‐mount DPDT slide switch rated for 4A/125VAC and 2A/250VAC.  
The  slide  protrudes  3/8"  from  the  mounting  surface  and  the  mounting  holes  are  spaced  1‐1/8"  on  center.    There  are 
openings  on  both  ends  of  this  switch  through  which  control/contact  cleaner  may  be  sprayed  while  the  slider  is 
positioned at the opposite end (with power removed from the equipment).  The low‐voltage position of this switch is 
rarely  needed,  and  if  a  compatible  replacement  cannot  be  found,  the  switch  could  simply  be  bypassed  in  order  to 
hardwire the microscope to operate at its nominal operating voltage.  Since only one pole on this switch is used, a SPDT 
replacement could be used. 
 
Figure 12 – Voltage Selector Switch Mounted to BHT/BHTU Electrical Base (100/115V Version) 
Olympus BH‐2 (BHT/BHTU) Electronics 
 
 
   Revision 2 
   
 
 
        Page 12 of 24