Olympus CH20 수리 매뉴얼 - 페이지 26

{카테고리_이름} Olympus CH20에 대한 수리 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Olympus CH20 31 페이지. Biological microscope

Olympus CH20 수리 매뉴얼
CH10/CH20

1-4 Voltage adjustment

ž When variable resistance (KR431400) and transformer (DL150400, DL150500) are replaced, it is
necessary to adjust the minimum voltage. The adjustment procedures are as follows:
ž Setup
Tools required: digital multimeter
Figure-B
Figure-C (100V system)
D. REPAIR PROCEDURE
Figure-A
i. Turn on the power, and fully turn the
VARIABLE RESISTANCE (KR431400) •
counterclockwise so that the brightness of
lamp can be minimum. (Figure-A)
ii.Measure AC voltage at the terminals, ‡ and
of the TRANSFORMER‚. (Figure-B)
(In 100V system, measure it after selecting
the frequency of slide switch S101.)
iii. In 100V system, adjust the voltage with R107
trimmer for 50 HZ, with R109 trimmer for 60
HZ, and in 200V system, with R107 trimmer
so that the minimum voltage can meet the
standard. (Figure-C, D)
Standard
Figure-D (200V system)
D-2
1.4 - 1.5V
±
,
1 2