Simplified Transmitted Light Observation
Frosted filter
Mirror
Base
3
Caution
· Do not project the image of the external light source filament on the frosted surface of the frosted filter.
Otherwise, the frosted filter may deteriorate.
· Use neutral detergent to clean the frosted filter.
· In transmitted light observation at a total magnification of no more than 10X, the field of view may be obscured
in the peripheral sections depending on the stage in use.
30
Downloaded from
ManualsNet.com
Microscope
body
Objective
Stage
External
light source
search engine
1
Illuminate the specimen with an external light source (LSD illuminator,
light guide illuminator, etc.).
Light the external light source as shown in the figure on the left and
irradiate the mirror assembly.
2
Eliminate irregularities in illumination.
1) Align the microscope body center with the center of the SZX-
STAD2 adapter.
2) Set the zooming knob on the microscope body to the minimum
magnification and focus the stage top surface.
3) While observing through the eyepiece, rotate the mirror knob to
adjust the mirror angle so that the entire field of view is illuminated
uniformly.
When observing using an oblique illuminator, remove the
TIP
frosted filter and, while observing through the eyepiece, tilt the
mirror gradually until optimum contrast can be obtained.