3. Sample Analysis Guidelines

3.1.
Handling Samples
Keep the sample chamber dry. Do notspill liquid into the sample chamber.
If a spill occurs, immediately invert the unit to remove the liquid.
Allow to fully dry before proceeding.
Condensation on the outside of test tube may result in read errors. Ensure that the
sample temperature has reached ambient temperature before placing the tube into the
sample chamber.
Never reuse test tubes for multiple samples
Bubbles or aeration within the sample may result in read errors. To remove air bubbles
from the sample, hold vertically and rapidly tap the outside of the tube with fingers.
3.2.
Positioning Samples
The orientation of the test tube may result in slightly different readings, especially for low
fluorescence readings. This is due to inconsistency in quality of the polymer test tube. For
proper use, ensure that the tab of the tube is oriented the same way for every measurement.
AccuStrata Fluorotek 200 User's Manual