Alemite 8424 서비스 매뉴얼 - 페이지 5

{카테고리_이름} Alemite 8424에 대한 서비스 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Alemite 8424 9 페이지. Flow meter

Flow Meter
The word "user"
indicates a
calibration factor set
by the user is being
used.
The flow chart
alongside shows the
switch over logic
from one display
page to another. In
this condition, the
Reset key permits
switching from User
factor to Factory
factor. To confirm the
choice of calibration factor, quickly press CAL while
"User" or "Fact" are displayed. After the restart cycle, the
meter uses the calibration factor that has just been
confirmed.
WARNING
When the Factory Factor is confirmed, the old User
factor is deleted from the memory.
In-field Calibration
This procedure calls for the fluid to be dispensed into a
graduated sample container in real operating conditions
(flow rate, viscosity, etc.) requiring maximum precision.
WARNING
For correct 8424 calibration, it is most important
to:
• completely eliminate air from the system before
calibrating;
• use a Sample Container with a capacity of not
less than 5 liters featuring an accurate graduated
indicator.
• ensure calibration dispensing is done at a con-
stant flow rate equivalent to that of normal use,
until the container is full;
• not reduce the flow rate to reach the graduated
area of the container during the final dispensing
stage (the correct method during the final stages
of sample container filling consists in making
short top-ups at normal operation flow rate);
• after dispensing, wait a few minutes to make sure
any air bubbles are eliminated from the sample
container; only read the Real value at the end of
this stage, during which the level in the container
could drop.
Alemite, LLC
• if necessary, carefully follow the procedure indi-
cated below.
In-field Calibration
Procedure
1. NONE - 8424 In Stand By
2. LONG CAL KEY KEYING - 8424 enters calibration
mode, shows "CAL" and displays the calibration
factor in use instead of total. The
words "Fact" and "USER" indicate
which of the two factors is currently
in use.
3. LONG RESET KEY KEYING -
8424 shows "CAL" and the partial at
zero. K24 is ready to perform on-site
calibration.
4. DISPENSING INTO SAMPLE CONTAINER -
Without pressing any KEY, start dispensing into the
sample container. Dispensing can be interrupted and
started again at will. Continue
dispensing until the level of the fluid
in the sample container has reached
the graduated area. There is no need
to reach a preset quantity.
5. SHORT RESET KEY KEYING -
8424 is informed that the calibration
dispensing operation is finished.
Make sure dispensing is correctly finished before
performing this operation. To calibrate the 8424, the
value indicated by the partial totalizer (example 9.800)
must be forced to the real value marked on the
graduated sample container. In the bottom left part of
the display an arrow appears (upwards and
downwards), THAT SHOWS the direction (increase or
decrease) of the USER K FACTOR value change when
the operations 6 or 7 are performed.
6. SHORT RESET KEY KEYING -
Arrow direction changes. The
operation can be repeated IF
NECESSARY.
7. SHORT/LONG CAL KEY KEYING -
The indicated value changes in the
direction indicated by the arrow - one
unit for every short CAL key keying
continually if the CAL key is kept pressed.
the first 5 units slowly and then quickly).
desired value is exceeded, repeat the operations from
point (6).
5
SER 8424
-
(for
If the
Revision (1-11)