Abbott i-STAT 1 사용자 설명서

{카테고리_이름} Abbott i-STAT 1에 대한 사용자 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Abbott i-STAT 1 9 페이지. For use with a clia certificate of waiver
Abbott i-STAT 1에 대해서도 마찬가지입니다: 빠른 참조 매뉴얼 (7 페이지)

Abbott i-STAT 1 사용자 설명서

Quality Check Messages and Codes

Message
Cause
Analyzer Error, Use
Handheld detects
Electronic
problem from which it
Simulator
is likely to recover.
Analyzer Error,
Handheld detects
See Manual
problem from which it
may not recover.
No display
Disposable batteries
dead or rechargeable
battery fully
discharged. Keypad
not responding. Start
switch broken.
Cartridge Locked
Dead battery(s).
does not disappear
Mechanical
after test cycle
problem.
completed
(01)00054749002139
Abbott Point of Care Inc.
100 and 200 Abbott Park Road
Abbott Park, IL 60064 • USA
Emergo Europe
Molenstraat 15
2513 BH The Hague
The Netherlands
Art: 714254-01S
(continued)
Action
Insert the Electronic Simulator. If
PASS, continue to use handheld.
Insert Electronic Simulator. If PASS,
insert a cartridge with sample
or control. If the code does not
reappear, continue to use handheld.
Change disposable batteries or
recharge battery. If still no display,
call Support Services.
Wait until handheld turns off. Turn
handheld on. If resets, remove
cartridge. If not, change or recharge
battery(s) and turn handheld on.
06F16 57
For information related to Article 33
of the EU REACH regulation
(EC No. 1907/2006), please refer
to PMIS.abbott.com. If you have
issues logging into the website,
contact Abbott at:
Rev. Date: 13-Feb-17
i-STAT
User Guide
.......................
©2017 Abbott Point of Care Inc. All rights reserved.
Printed in USA. i-STAT is a registered trademark of the
Abbott Group of Companies in various jurisdictions.
i-STAT
®
1
®