Note: Intensity and trends may vary. The important thing is the signal starts increasing.
Step 7) Perform Flask Transfer
Once the Tails are identified, perform a flask transfer as per usual. Keep in mind that the signal will be very
low and all ambient lighting effects will greatly increase
2018-2021 Fraction Finder
Step 8) Turn Off Recording
will turn red. Then, plug in a USB stick to the monitor. Press "Transfer" to transfer your run data to a USB
stick. It will upload a file in CSV format, easy to work with in Excel.
2022 and Later Fraction Finders
Step 8) Turn Off Recording
The Record button will turn red. Then, remove the sd card and transfer the data via the SD card reader.
Files will upload in CSV format, easy to work with in Excel.
Section 7: Issues
1) If the signal is an abnormally low signal, a few things should be checked:
a. Ensure that the alignment of the FRACTION FINDER sensor is correct. This can be done by
trying to align the screw in the "light shield" and the top of the light shield with the cooling
water port of the condenser; if they are aligned, the alignment is optimal. The "light
shield" is the half-circle piece of hardware adjacent to the light source.
b. Ensure that the integration time is set correctly by using the AID (Auto Integration) AUTO in
Settings. The checkbox should be checked off.
c. Ensure that the Scans to Average is set to 5.
d. Are you distilling THC? Our THC Indicator is a short, broad signal; however, it is still easily
discernible. See our Chemical Cheat Sheet.
e. If none of the above seem to be the origin of the problem, please take a photo of both
your spectrum plot and your wavelength plot, then email [email protected].
Arometrix, Inc.
(optional) If you record your run, tap "Record Off". The circle above this button
(optional) If you recorded your run with an sd card installed, tap "Record Off".
Optimal Sensor and Light-Blocking Tape Orientation
arometrix.com/resources