10. Checklist for safety- and function test
Safetytest
Enclosure Leakage Current
Or
Enclosure Leakage Current

Functiontest

1. Turn the power ON at the ARTROMOT® -K3
Display: Software Version K3 V40 161002
2. The set range of motion for Extension/Flexion is from -5 to 110 degrees.
Verify the angle in a position of 0 degrees with a tolerance of +/- 5 degrees.
Verify the angle in a position of 60 degrees with a tolerance of +/- 5 degrees.
Verify the angle in a position of 105 degrees with a tolerance of +/- 5 degrees.
3. Safety - OFF Functions.
Start the ARTROMOT® -K3 in a continuos mode.
Press any of the buttons to stop the motor immediately.
Verify this for all buttons.
4. Perform the following settings.
Extension:
Flexion:
Pause Ex.:
Pause Flex.:
Force:
Speed:
Adjust a maximum femur length.
5. First press "STOP" and afterwards press "START".
The ARTROMOT®-K3 should then reach both points of changing direction
within 60 to 75 seconds.
6. Change the speed to 50% and afterwards press "START".
The ARTROMOT®-K3 should then reach both points of changing direction
within 105 to 120 seconds.
7. Verify the settings for "Pause Ex." and "Pause Flex."
The tolerance for both is +/- 3 seconds.
EN60601-1/ VDE 0751
UL2601 (115 Volt)
and press "SET" .
-5 degrees
110 degrees
5 sec.
10 sec.
45 kp
100%
Measuring
100 µA


100 µA
12
Date/Signature
µA
µA
OK
Error