Browse online or download pdf User Manual for Medical Equipment Amca MiKD-350-S. Amca MiKD-350-S 20 pages. Hand-desinfectiontower with fiebermessfunktion
2.First Use
2.2. First use
2.2.1. Remove all packaging and label materials and dispose of them according to
applicable local regulations.
2.2.2. lnstall the device in accordance with the Installation Manual.
2.2.3. Ensure !hat the device is in an upright position on a stable,
dry,
and non-slip
2.3. Preparation before first use
2.3.1. Open the cover on the disinfectant tank (carboy) and fill it with an appropriate
disinfectant.
2.3.2. Connect the power plug of your device (AC 220 V).
2.3.3. After the LED indicating that the device is operating (activated) lights
contact is made several times to ensure !hat the disinfectant of the device
reaches the nozzles (the air escapes). During this process you may hear a
&
different sound from the
NOTICE: Ensure that your device works with the help of the sensor until
disinfectant reaches the nozzles of the device. lt is normal to hear a different
sound from the pump during this
2.4. INST RUCTIONS FOR USE OF T HE MiDiSiS HAND DISINFECT ION DEVICE
WITH THERMOMET ER
2.4.1. Operating Conditions of the Device:
2.4.1.1.
This device is designed for indoor use.
2.4.1.2.
lt cannot measure accurately if exposed to direct sunlight.
2.4.1.3.
lt cannot measure accurately in a place with air flow (do not place it close
to air curtains at the entrances in shopping malls etc.).
2.4.1.4.
Measurements with glasses are not
2.4.1.5.
The forehead should be open (if the forehead is not seen directly due to
forelock, hat,
correctly).
2.4.1.6.
lt cannot measure correctly if the forehead is sweaty.
2.4.1.7.
The forehead should be positioned centrally to the target.
2.4.1.8.
lf the device is mov ed from its environment to an environment with a
different
to the new environm ent.
2.4.1.9.
The ambient temperature sensor on the back of the device should hang
out of the housing.
2.4.1.10. The device can measure accurately at ambient temperatures between 15-35 °C.
2.4.1.11. Measurement accuracy is + -0.5 °C under right conditions.
2.4.1.12. Do not blow hot air into the device.
2.4.1.13. Do not wipe the temperature sensor when cleaning the device. Do not bring
the sensor into contact with water or liquid cleaning agents.
2.4.1.14. The device is calibrated for the human skin. lt cannot accurately measure
external objects.
surface.
pump,
which is normal.
long hair in front of
temperature,
it requires an average of 30 minutes to get adapted
procedure.
healthy.
head, etc.,
the device cannot measure
smooth, clean,
up,
sensor