3 SYSTEM SETUP
3.1 Measurement Location
1)
Choose a location near to the entrance of the area you want to protect.
2)
There must be space to install the VIRALERT 3 LITE so that the Measurement
Subject can face the system from a distance of 1 m / 3 ft.
3)
Avoid locations where there are possible heat sources e.g. electrical
equipment, radiators etc. in the field of view of the Imager.
4)
Also avoid locations where the Imager may pick up reflections from windows,
mirrors etc.
5)
VIRALERT 3 LITE is designed to scan one person at a time. The measurement
location should not be a high-traffic area where people other than the subject
will be in the field of view.
6)
During the temperature measurement process, the Subject will need to stand
still in front of the Imager for approximately 2 seconds. Make sure there is
adequate provision for social distancing in your chosen measurement location.
7)
The Imager must have a clear, unobstructed view of the Subject's face, ideally
with the Imager at a height around eye level. If necessary, use one of the
mounting accessories available from AMETEK Land (see Section 1.3).
8)
The Blackbody Reference Source requires a mains voltage supply. You will
also need to provide power to the Laptop/PC running the VIRALERT 3 LITE
software.
9)
Once the system is installed and operating, use cable ties to secure the
cables to suitable structures. This minimises the risk of the cables becoming
dislodged and reduces the risk of tripping over the cables.
10) Posters which explain the screening process and provide instructions for your
staff and visitors on how to use the system are available for download from:
www.landviralert.com
Detailed guidelines for deployment, implementation and operation of a
screening system can be found in ISO/TR 13154:2017
3.2 System Connections
NOTE
If you have purchased a mounting accessory i.e. Surface Mount Bracket, Clamp
Mount Bracket, or Tripod from AMETEK Land, then each accessory will come supplied
with its own installation instructions.
04
QUICK START GUIDE
HUMAN BODY TEMPERATURE SCREENING SYSTEM