result: ascorbic acid, uric acid, hemoglobin,acetaminophen,
Dopamine, L-dopa, L-dopa and Tolbutamide etc. These sub-
stances do not affect test results in normal concentration but
may affect test results in high concentration. Do not use hae-
molysis sample, icterus sample or high lipemia samples.
※ The following compounds, when determined to in excess of
their limitation, may produce inaccurate results:
Compound
Acetaminophen
Ascorbic Acid
Bilirubin
Cholesterol
Creatinine
Dopamine
EDTA
Galactose
Gentisic Acid
Glutathione
Haemoglobin
Heparin
Lbuprofen
Lcodextrin
L-dopa
Maltose
Methyldopa
Pralidoxime Iodide
Salicylate
Tolazamide
Tolbutamide
Triglycerides
Uric Acid
Xylose
29
Limitation
≤8.0mg/dL
≤5.0mg/dL
≤90mg/dL
≤500mg/dL
≤5.0mg/dL
≤2.0mg/dL
≤360mg/dL
≤900mg/dL
≤5.0mg/dL
≤53mg/dL
≤500mg/dL
≤8000U/dL
≤50mg/dL
≤13mg/dL
≤10mg/dL
≤900mg/dL
≤3.0mg/dL
≤25mg/dL
≤60mg/dL
≤100mg/dL
≤400mg/dL
≤2000mg/dL
≤8.0mg/dL
≤100mg/dL
※ Patients undergoing oxygen therapy may yield falsely lower
results.
※ Not for use for patients in a hyperglycemic-hyperosmolar
state, with or without ketosis.
※ Not for use on critically ill patients.
※ Not to be used for patients who are dehydrated, hypertensive,
hypotensive or in shock.
※ Very low (less than 20%) or very high (more than 60%) red
blood cell count (hematocrit) can lead to incorrect test results.
If you do not know your hemarocrit level, please consult your
health care provider.
DISPLAY MESSAGE
MESSAGE
WHAT IT MEANS
Blood glucose level is
lower than 20mg/dL
(1.1mmol/L)
Blood glucose level is
higher than 600mg/dL
(33.3mmol/L)
Meter is ready to test
control solution under
the CTL state.
Battery power is low.
ACTION
The message indicates
very low blood sugar. You
should consult with your
healthcare professional.
This indicates sever
hyperglycemia (high
blood glucose).
You should seek imme-
diate medical assistance.
Please place a drop of
control solution onto the
test well of test strip.
Please change the
battery.