This symbol means: "Consult operating instructions". To achieve accurate results, read the
instructions for use carefully before use and follow all instructions provided therein. Keep the
instructions for use for future reference.
Your ADE electronic scale is a quality, sturdy and durable product manufactured under the high standards and
regulations for calibrated class III scales in accordance with national regulation in Germany. The scale
signed and manufactured for the daily practice in hospitals, doctors' surgeries and in-patient nursing facilities.
The scale is especially being designed for the use of adipose patients.
ADE instruction manuals have been prepared to provide the user with all the necessary knowledge about the
safe use of ADE scales and their related accessory equipment.
For further questions about how to use the products or about this or other service manuals, feel free to contact
your local ADE representative or visit our web page at www.ade-hamburg.de
The indicator/ display of your ADE scale is able to be mounted on walls, on sockets, handrails or on stand-on
devices for the table depending on the model you have selected.
Please insure that your scale is being placed vibration free of any vibration and at a dry place.
Please take the indicator and the cable extension out of the packaging and follow the instruction for assembly.

Attachment of the indicator

The ADE scales are delivered in a largely assembled state. Only model M301020 requires the indicator to be
connected to the weighing platform. Connect the indicator to the scale platform using the connector provided.
Alignment of the scale
Place the scale on a plain, dry and level surface.
Adjust the scales so that all feet are in contact with the surface.
The air bubble in the spirit level should be right in the centre of the circle
If the scale has been transported or stored, it must be re-adjusted as described above.

Power supply

Power is supplied either by the rechargeable battery in the display unit or by the mains power via the supplied
power unit. The imprinted voltage value at the mains adapter must match the local mains voltage.
After checking same connect the mains cable to the indicator and to a socket.
Note: The wall socket must be located near to the unit and be easily accessible. Please use only the mains
adapter supplied by ADE. The mains adapter is charging the battery pack at the same time.
When the battery is being charged for the first time a sufficient charging period of approx. 24 hours is neces-
sary to achieve a full battery load.
In case the scale is being used with batteries only and the display indicates – Lo, the battery should be re-
charged with the ADE mains adapter. After the first charging period the battery need 12-16 hours. The battery
should be charged while the scale is switched off. Upon completion of the charging period switch the scale on
again to check the battery status. In case Lo- still appears on the display inform your service partner.