Alignment of the scale
1) Place the scale on a plain, dry and level surface
2) Undo the locknuts of the four feet of the scale and unscrew them until they got contact to the surface
3) The air bubble in the spirit level must be right in the centre of the circle
4) Please ensure that the scale has no other contact to the floor than by the four feet
4) For all scales with a foldable column or a foldable Handrail:
(M500020-03; M500020-E-01; M500020-02; M500020-A-02; M500020-B-01; M500020-C-01; and M500020-04)
) Release the handle (pic 4a) at the bottom of the column or the handrail.
a
b) Upright the column or the handrail.
c) Finally re-tighten the handle (pic 4b) at the bottom of the column or the handrail again.
(Pic 4a)
Please ensure that the foldable handrail and column of the scale is fixed and stable in order to
avoid any unexpected moving of same. After transport or movement of the scale follow step 1 to 4
again.
Power supply
Power is supplied either by the rechargeable battery in the display unit or by the mains power via the supplied
power unit. The imprinted voltage value at the mains adapter must match the local mains voltage. Connect the
mains cable to the indicator and to a socket.
The wall socket must be located near to the unit and be easily accessible. Please use only the mains adapter
supplied by ADE. The mains adapter is charging the battery pack at the same time.
When the battery is being charged for the first time a sufficient charging period of approx. 24 hours is neces-
sary to achieve a full battery load.
If the unit is in battery mode and the display indicates – Lo, the battery needed to be recharged with the mains
adapter. For a full battery load a charging period of 12-16 hours is required with the scale being switched off.
If–Lo still appears at the display inform your service partner.
Replacing rechargeable batteries
The battery compartment is located on the front of the unit.
1) Open the battery compartment.
2) Remove the battery cables carefully from the housing, connect the wires (red to red, black to black) and
insert the battery into the housing.
3) Close the battery compartment again.
Only use original rechargeable batteries and charger units from ADE. The warranty does not
cover any damage caused by using other types of batteries or charger units.
M500020-210322-Rev009-UM-e