Automess 6150AD Technical Manual - Page 4
Browse online or download pdf Technical Manual for Cables and connectors Automess 6150AD. Automess 6150AD 7 pages. Dose rate meter probe connector
06/2003
A typical setup to read the 6150AD by a PC (optionally with a probe connected) would be:
standard probe cable
probe (optional)
The cable EB84-0193 is basically a probe extension cable with a »T« connection making the GND and
Term signals available at the 9 pin sub-D connector.
2.1.2 Software Characteristics
Interface settings are
4800 Bd (9600 Bd for special 6150AD1-BiZa version)
8 data bits
no parity
1 stop bit
The character string sent approximately every second consists of the following six bytes:
1. STX (02H) as the start of text character
2. type of detector/6150AD
3. dose rate mantissa (low byte)
4. dose rate mantissa (high byte)
5. dose rate exponent
6. block check character
The second byte (type of detector/6150AD) contains:
Bit 0-5:
detector in use:
Bit 6:
type of internal tube: 0 = ZP1200 (6150AD2/4/6)
Bit 7:
measuring quantity:
Daimlerstrasse 27
D-68526 Ladenburg
Technical Manual 6150AD Probe Connector
connecting cable »PC - 6150AD - probe«
Automess part number EB84-0193
0 = probe 6150AD-0
7 = probe 6150AD-b
15 = probe 6150AD-15
17 = probe 6150AD-17
18 = probe 6150AD-18
19 = probe 6150AD-19
20 = 6150AD internal tube
21 = probe 6150AD-t, low range tube
22 = probe 6150AD-t, high range tube
1 = ZP1310 (6150AD1/3/5)
1 = 6150AD »/E« model (6150ADx/E)
0 = other (6150ADx or 6150ADx/H)
Telefon +49(0)6203-9503-00 http://www.automess.de
Telefax +49(0)6203-9503-29 eMail: [email protected]
DB-9
PC
6150AD
USt.-ID-Nr. / VAT ID No.
DE 144 458 420
Page 4 of 14