Guidance and manufacturer's declaration - electromagnetic immunity
The BP3L is intended for use in the electromagnetic environment specified below. The customer or the user
of the BP3L should assure that it is used in such an environment.
IMMUNITY test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
For all ME EQUIPMENT and ME SYSTEMS
IEC 60601test level
± 6 kV contact
± 8 kV air
3 A/m
Table 2
Compliance level
± 6 kV contact
± 8 kV air
3 A/m
15
Electromagnetic environment -
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30 %.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.