b )The application of the cuff on any limb with intravascular access or therapy, or an arteriovenous (A-V) shunt.
c )The application of the cuff on the arm on the side of a mastectomy.
d )Simultaneous use with other medical monitoring equipment on the same limb.
e ) The blood circulation of the user needs to be checked.
3. Do not use this product in a moving vehicle as this may result in inaccurate measurements.
4. Blood pressure measurements determined by this product are equivalent to those obtained by professional
healthcare practitioners using the cuff/stethoscope auscultation method within the limits prescribed by the
American National Standard, Electronic or Automated Sphygmomanometer.
5. If you are using a smart phone to operate the device and a phone call comes in during the measurement,
the measurement process will be terminated automatically. It is thus recommended that the phone be set
in Airplane mode during measurement to avoid interrupting the measurement.
6. If an Irregular Heartbeat (IHB) is detected during the measurement procedure, the IHB symbol will be
displayed. Under this condition, the Wireless Blood Pressure Monitor can keep functioning, but the results
may be inaccurate. Please consult your physician for accurate assessment.
The IHB symbol will be displayed under 2 sets of circumstances:
1) The coefficient of variation (CV) of pulse period >25%.
2) The difference of adjacent pulse period is ≥0.14s and more than 53 percent of the total number of pulses
readings falls within this definition.
7. Please do not use any cuff other than that supplied by the manufacturer as this may result in inaccurate
measurements.
8. For information regarding potential electromagnetic or other interference between the Wireless Blood
Pressure Monitor and other devices together with advice regarding avoidance of such interference, please
see ELECTROMAGNETIC COMPATIBILITY INFORMATION. It is suggested that the Wireless Blood Pressure
Monitor should be operated at least 10 meters away from electric or wireless devices (e.g. routers,
4