iHealth View User Manual - Page 10

Browse online or download pdf User Manual for Blood Pressure Monitor iHealth View. iHealth View 12 pages. Wireless blood pressure wrist monitor

iHealth View User Manual
ELECTROMAGNETIC COMPATIBILITY INFORMATION
BP7S is intended for use in the electromagnetic environment specified below. The user of BP7S should ensure that it is used in such an
environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
BP7S is intended for use in the electromagnetic environment specified below. The user of BP7S should ensure that it is used in such an
environment.
IMMUNITY test
Electrostatic
discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: U
is the a.c. main voltage prior to application of the test level.
T
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture's declaration - electromagnetic emissions
Compliance
Group 1
Class B
Class A
Complies
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer's declaration - electromagnetic immunity
IEC 60601test level
± 6 kV contact
± 8 kV air
± 2 kV for power supply
lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
<5 % U
T
(>95 % dip in U
)
T
for 0.5 cycle
40 % U
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
<5 % U
T
(>95 % dip in U
)
T
for 5 s
3 A/m
Table 1
Electromagnetic environment - guidance
BP7S uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
BP7S is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes.
Table 2
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power supply
lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
<5 % U
T
(>95 % dip in U
)
T
for 0.5 cycle
40 % U
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
<5 % U
T
(>95 % dip in U
)
T
for 5 s
3 A/m
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30 %.
Main power quality should be that of a typical
commercial or hospital environment.
Main power quality should be that of a typical
commercial or hospital environment.
Main power quality should be that of a typical
commercial or hospital environment. If the user of
BP7S requires continued operation during power
main interruptions, it is recommended that BP7S
be powered from an uninterruptible power supply
or a battery.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.