ARTROMOT K4 Service Manual - Page 14
Browse online or download pdf Service Manual for Fitness Equipment ARTROMOT K4. ARTROMOT K4 18 pages.
9. Special function Service Menu
Function of service Menu
Calibration
Display contrast
Error log
Device runtime
Entering the service menu:
Press the menu key until Service Menu
shows up (menu 3).
Press
for
is flashing on the display.
The display will change and show:
Entering code.
For the code press: 1 3 2 4
Now you see the symbols of the service menu
for 5 seconds,
9.1 Calibration
ATTENTION!
Before you do a calibration switch the device
OFF and ON.
Adjust a maximum femur length, a minimum lower leg
length, hip axis in smallest position and the middle
position of the foot rotation.
Press the symbol calibration
Display:
Press + or - to move the ARTROMOT®-K4 to 0
degrees.
Press START, the calibration starts automatically. The
device will reach both maximum points and move
between –10 bis 125 degrees with different speed.
Wait until the ARTROMOT®-K4 stops.
If the calibration was succesful the device stops at
0 degrees and show following symbols on the display:
Display:
Press STOP twice to leave the service menu.
Finally, a safety and function test has to be performed
(see chapter 11).
9.2 Display contrast
Press the symbol display contrast
Display:
Press + or - to set up the requested display contrast.
You can set the display contrast from 0 – 100%.
Press STOP twice to save the settings and leave the
service menu.
14
.
.
XX %