Follett FZR5-PL Series Installatie-, Bedienings- en Onderhoudshandleiding - Pagina 11
Blader online of download pdf Installatie-, Bedienings- en Onderhoudshandleiding voor {categorie_naam} Follett FZR5-PL Series. Follett FZR5-PL Series 17 pagina's. -30 c undercounter plasma freezer
Ook voor Follett FZR5-PL Series: Installatie-, Bedienings- en Onderhoudshandleiding (20 pagina's), Installatie-, Bedienings- en Onderhoudshandleiding (20 pagina's)
Refrigeration system
The FZR5-PL -30 C (-22 F) freezer refrigeration system is designed to give many years of trouble-free service.
Except for routine cleaning of the air-cooled condenser and related parts, the refrigeration system requires no
service or maintenance. The system uses a thermostatic expansion valve and is critically charged. Access fi ttings
are provided for ease of service. However, the connection of refrigeration service hoses to the fi ttings will almost
invariably result in a signifi cant change in the system charge. This change can adversely affect the performance
of your freezer. Therefore, Follett recommends that if hoses are ever connected to the refrigeration system for
service, the refrigerant should be recovered, the system evacuated, and recharged by weighing in the correct
refrigerant charge.
Note: Do not charge the system by pressures.
Checking refrigeration system pressures
1. Remove the rear access panel (Fig. 12).
2. Turn the power switch to the on position.
3. Following the instructions on page 5 verify that the temperature controller is set to the original factory
set point cut out setting of -35 C (-31 F).
4. Allow the freezer to operate and stabilize at least 60 minutes, verifying the cut-out temperature is
being reached.
5. Connect refrigerant hoses to access fi ttings and measure air temperature at condenser intake grille.
6. Verify correct pressures with the temperature chart below.
7. Troubleshoot refrigeration system as needed.
Note: Do not attempt to obtain correct refrigeration pressures by adjusting the system charge.
Condenser inlet air temperature
Discharge pressure (psi)
Suction pressure (psi)
Refrigeration system diagram
Low Psi
Vapor
Low Psi
Liquid
EVAPORATOR/DEFROST UNIT
HIGH PRESSURE VAPOR
70 F
197
4
HEAT EXCHANGE
TXV
HIGH PRESSURE LIQUID
80 F
225
5
LOW SIDE
SERVICE PORT
SERVICE PORT
DRYER
LOW PRESSURE LIQUID
11
COMPRESSOR
10oz. Charge
High Psi
Vapor
HIGH SIDE
High Psi
Liquid
CONDENSING UNIT
LOW PRESSURE VAPOR