Andon AG-605A Bedieningshandleiding - Pagina 6
Blader online of download pdf Bedieningshandleiding voor {categorie_naam} Andon AG-605A. Andon AG-605A 16 pagina's.
of temperature will appear. Short press and release the MEM button until
the unit (℃ or ℉) that you prefer appears.
9. Short press the S button to choose whether to delete the memory. Short
press and release the MEM button to delete the memory and the meter will
automatically power off. If you don't want to delete the memory, short press
the S button directly to turn off the meter.
10. When you change the batteries, you have to readjust device parameter.
Coding the Meter
Code meter before using it for the first time and every time you change
another vial of test strip. With the meter turned off, insert the code card and
pull out after the OK appears on screen.
Before code the meter, make sure the test strip have not expired.
Testing Your Blood Glucose Level
Before testing, you will need the following items:
Glucose AG-605A Meter
AGS-1000 test strip
Adjustable Automatic Lancing Device
Sterile Lancet
CAUTION:
To reduce the chances of infection:
Choose a clean, dry work surface.
Never share a lancing device and lancet with another person.
Always use a new and sterile lancet. Lancets are for single use only.
Always use a new Test Strip. Test Strips are for single use only.
Avoid getting lotion, oils, dirt or debris in or on the lancet and lancing device.
Step 1
Wash hands using soap and warm water. Rinse and dry thoroughly.
Step 2
Make sure you have coded the meter and loaded the needle into the lancing
device.
Step 3
Remove a Test Strip from the vial and replace the cap on the vial immediately.
Step 4
Insert the test strip into the test strip slot of the meter. The meter turns on
automatically, and the code number appears on the screen.
6