Attonic AD-TM50 Bedieningshandleiding - Pagina 10
Blader online of download pdf Bedieningshandleiding voor {categorie_naam} Attonic AD-TM50. Attonic AD-TM50 20 pagina's.
7-3.Set comparator HI (Initial value: 0000)
〇 TRACK / PEAK Mode
・You can change the HI side setting value of the comparator.
A comparator is a comparison with a set value.
For example, if you set the point of high value (HI) to 1.00N・m and the point of low value (LO) to 0.50N・m
and gradually apply force from 0.00N・m, LO will be indicated with a long intermittent sound at first,
but the display and buzzer sound will disappear when the force exceeds 0.50N・m.
When the force exceeds 1.00N・m, HI will be indicated and a short intermittent sound will be heard.
* When the HI or LO value is set (non-zero), the external output is also outputted in conjunction with it.
* The setting value should be entered as an absolute value (regardless of whether it is positive or negative).
This digit blinks.
* It returns to the beginning when the "DAT" switch is pressed 4 times.
The value will increase if the
"POW" switch is pressed
while this digit is blinking.
〇 PP Mode
・You can change the setting value on the + side (upper section) of comparator.
For example, if you set the point of + side (HI) to 1.00N・m and the point of - side (LO) to -0.50N・m, and
gradually apply force to the + side from 0.00N・m, the display of the upper section will be changed, and
when the force exceeds 1.00N・m, a short intermittent sound will be heard.
* The setting value should be entered as the value of + side instead of an absolute value.
* A short intermittent sound will be heard if the HI setting value (+ side) is exceeded, and a long one if the LO
setting value (- side) is exceeded. After that, even if the HI peak value is exceeded, a long intermittent sound
continues.
* A long intermittent sound will be heard if the LO setting value (- side) is exceeded, and a short one if the HI
setting value (+ side) is exceeded. After that, even if the LO peak value is exceeded, there is still a short
intermittent sound.
* The position of the decimal point varies with each unit and each model.
* For models that increase in digits, the maximum setting value is 9999.
For example: When the unit is N・m in AD-TM100, the settable maximum value is 9.999(N・m).
* The setting value returns 0 when the mode is changed.
* Short press the "FUN" switch to confirm the setting. Next is the setting for comparator LO.
The digit shifts to the
"DAT" Switch
left by one place
every time the
"DAT" switch is
pressed.
"POW" Switch
"DAT" Switch
This digit blinks when
the "DAT" switch is
pressed 3 times.
"POW" Switch
"POW" Switch
"9" is indicated when the "POW"
switch is pressed 9 times.
It returns "0" when the "POW" switch
is pressed again.
"DAT" Switch