ADE M500020-D Bedieningshandleiding - Pagina 10
Blader online of download pdf Bedieningshandleiding voor {categorie_naam} ADE M500020-D. ADE M500020-D 16 pagina's. Electronic wheelchair and multipurpose scales
Tare function
The TARE key (9) is used when the patient is being weighed with other items (e.g. wheelchairs, clothing,
shoes or other items) which should be ignored by the scale. To do this, weigh these items first. Weigh these
items first. After the weight is displayed, press the TARE key (9) to reset the display back to zero. The display
changes to net mode (NET status indicator (5) is on) and shows zero.
Entering manually tare - values:
The user is able to insert manually the gross weight, which is for example mentioned at the identification plate
of the wheelchair. Enter the appropriate tare value just by using the key pad under normal weighing mode.
The value will be stored after use of the ENTER key (or TARE). The ADE-display is now ready for weighing
and changes to net mode [NET status indicator (5) is on]. Enter the tare button to leave this mode.
Preset tara
Press and hold the PRESET TARE key (10) for approx. 3 seconds. PRO PT X appears on the display. Select
the memory location 1-10. Now enter the TARE VALUE which you would like to be ignored by the scale and
confirm same with ENTER. The value is displayed as a negative figure and will be deducted when a load is
being placed on the scale. Enter the on/off, zero – button (8) to leave this mode.
Restore the preset Preset tara value:
Press the PRESET TARE key (10). "Pt x" appears on the display. Enter a number from 0-9 (memory location
for the tare value). If no new number will be inserted to change the already displayed value within five sec-
onds, the unit will turn back to weighing mode and displays the last stored value as negative figure as above.
Enter the on/off, zero – button (8) to leave this mode.
Displaying the gross weight, net weight and tare value:
Press the "0" key to change between the net weight (gross weight minus the tared value) and the gross
weight.
Hold function
The determined weight could be continued being displayed even when the load has already been removed.
That enables taking care of the patient first before reading the results.
To use the hold function the scale must be at standstill. The standstill status indicator
(14) is on and
Hold appears at the digital display. Deactivate the HOLD function by pressing the key again.
Determining the Body Mass Index (BMI)
The BMI figure is a index which displays the ratio between height and weight.
body weight
BMI=
(height)²
To determine this value you need the height of the person. Best way to get the height is of course to use an
ADE height measuring rod; model MZ10023.
Start the scale and weigh the person. Then press the BMI key (11) to start BMI mode. The last value entered
appears on the display. Use the numerical key pad to enter a new value for the height (in cm) or retain the
value which is already displayed.
Confirm the value entered with the ENTER key or delete same by using the CLEAR key. After the entry has
been confirmed by the ENTER key, the BMI value will be calculated using the weight on the scales. The dis-
play changes between weight value and BMI value "b xx.x" (every two seconds) for a period of 3 minutes.
Press the BMI key (11) to cancel before this time has expired.
10
M500020-210322-Rev009-UM-e