Accutrol T-ABVA Installatie-, bedienings- en onderhoudshandleiding - Pagina 10

Blader online of download pdf Installatie-, bedienings- en onderhoudshandleiding voor {categorie_naam} Accutrol T-ABVA. Accutrol T-ABVA 16 pagina's.

Installation
Step 3: Locate the Hood Probe
Due to the wide variety of fume hoods and configurations available,
there is no single hood probe location that will work for all fume
hoods. Contact the fume hood manufacturer to obtain the
optimum side wall probe location for the specific model hood
being used.
If no information is available, generally good performance is
achieved when the probe is mounted as shown in Figure 2.
Step 4: Install the Hood Probe
Before proceeding, reference the fume hood manufacturer
recommendations for best method and safety precautions for drilling
the fume hood sidewall.
a. Drill a 7/16" diameter hole in the side wall of the fume hood in
the location selected in Step 3.
b. Working from the inside of the fume hood, insert the end of the
tubing through the drilled hole feeding the entire length through
up to the probe.
Drilled Hole
(Inside of
Fume Hood)
Step 5: Wiring the Terminal Strip
On the back of the ABVA there is a 5-conductor (screw-less) terminal strip used for connecting the
"Remote Mute Input" and "Alarm Relay Output". If these features are not being used, proceed to Step 6.
Otherwise, route the wires through the appropriate holes or electrical box knock-outs and connect to the
respective terminals. Note: Connections to the terminal strip are made by pushing the button above the
terminal, inserting the stripped end (1/4") of the wire into the hole and releasing the button.
Step 6: Connecting Input Power
Route the cord end of the DC Power Module through the appropriate hole or electrical box knock-out and
insert into the power plug located on the back of the ABVA.
Step 7: Connect the Tubing to the ABVA
Route the hood probe tubing through the appropriate hole or electrical box knock-out, using caution not to
cause any kinks. Slide the tubing over the sensor port on the back side of the ABVA.
Step 8: Secure the ABVA
Secure the ABVA to the mounting surface using the appropriate two screws provided.
T-ABVA Manual |Rev.C, JULY 2017
c. Firmly insert the probe into the predrilled hole. It
should not be required; however you may apply some
sealant / adhesive to the back of the probe flange to
secure the probe to the sidewall if desired.
Sash 100% Open
Figure 2
Page 6