Arometrix Extraction Finder Gebruikershandleiding - Pagina 13
Blader online of download pdf Gebruikershandleiding voor {categorie_naam} Arometrix Extraction Finder. Arometrix Extraction Finder 14 pagina's.
Ook voor Arometrix Extraction Finder: Snelstarthandleiding (2 pagina's), Gebruikershandleiding (20 pagina's)
Pictured here is the Spectrum View, which tracks
molecules as a function of
Intensity over Wavelength.
Section 8: Issues
1) If the signal is an abnormally low signal, a few things should be checked:
a. Ensure that the EXTRACTION FINDER system is properly installed.
b. Ensure that the integration time is set correctly by using the AID (Auto Integration) AUTO in
Settings. The checkbox should be checked off.
c. Ensure that the Scans to Average is set to 5.
d. If none of the above seem to be the origin of the problem, please take a photo of both
your spectrum plot and your wavelength plot, then email [email protected].
2) If the signal looks sporadic and very abnormal, a few things should be checked:
a. Is there turbulent flow in the Extraction Adapter? Turbulent flow is not ideal. For the best
signal, ensure consistent, full flow. The addition of a small sight glass somewhere on top of
the Extraction Adapter might help you ensure this.
b. If the pressure is still changing quickly, wait for the pressure to become more constant.
c. Ensure that the integration time is set correctly by using the AID AUTO.
d. Increase the scans to average:
i. This should not be set significantly higher than 5 - keep the value below 15
e. Ensure that the Extraction Adapter is not in a location with a high amount of vibration
f.
Ensure that the Extraction Adapter is not in a location monitoring the vapor phase
g. If none of the above are the problem – ensure that liquid is still flowing
h. If none of the above seem to be the origin of the problem, please take a photo of both
your spectrum plot and your wavelength plot and contact customer service by emailing
3) If the system is not detecting the sensor (bottom left light on panel is red or yellow)
a. If the system was turned on, give the system up to 5 minutes, it may detect
b. Ensure that all cables are connected securely – especially the cable connecting the display
unit to the sensor
i. It may be easier just to disconnect and reconnect the cables from the display unit
and the sensor unit
c. Try identifying and using a new source of power to power the EXTRACTION FINDER display.
d. If none of the above seem to be the origin of the problem, please take a photo of both
your spectrum plot and your wavelength plot and email [email protected].
Arometrix, Inc.
Pictured here is the Wavelength View, which
or submitting the Support Request form on our website
tracks molecules as a function of
Intensity over Time.
arometrix.com/resources
13