5
Motor over current:
The current for the motor exceeded the maximum limit
->
Check the mechanics
->
Replace motor control (Pos. 1.35)
->
Replace motor (Pos. 3.1)
6
Motor control error:
Internal error in the motor control.
->
Replace motor
control (Pos. 1.35)
7
Eprom access error:
Memory error in the access of the
->
Replace hand held programming unit (Pos. 1.31)
8
CPM ROM error:
Memory error in the motor control
->
Replace motor
control (Pos. 1.35)
9
Communication:
Communication to the motor control is not possible
->
Check spiral cable and connector
->
Replace
hand held programming unit (Pos. 1.31)
->
Replace
motor
control (Pos. 1.35)
10
Unkown error in the motor control:
Unknown error in the motor control
->
Replace
motor
control (Pos. 1.35)
11
Motor enable timeout
Motor could not be enabled in time
->
Replace
motor
control (Pos. 1.35)
12
Invalid parameter motor error:
Motor has received a invalid parameter
->
Replace motor
control (Pos. 1.35)
->
Replace hand held programming unit (Pos. 1.31)
13
Stop release error:
The motor could not be released
->
Replace motor
control (Pos. 1.35)
->
Replace hand held programming unit (Pos. 1.31)
14
Unexpected motor Stop:
->
Check cables and connectors
->
Replace motor
control (Pos. 1.35)
15
Motor disabled:
Motor control disabled the motor.
->
Replace motor
control (Pos. 1.35)
16
Wrong command in the motor :
->
Replace motor
control (Pos. 1.35)
->
Replace hand held programming unit (Pos. 1.31)
17
5V supply error:
5V supply of motor control not sufficient
->
Replace motor
control (Pos. 1.35)
18
Initialise error real time clock:
->
Replace hand held programming unit (Pos. 1.31)
19
Communication error real time clock:
->
Replace hand held programming unit (Pos. 1.31)
20
Error real time clock:
->
Replace hand held programming unit (Pos. 1.31)
EPROM.
.
3
21
Range exceeded:
The measured angle is out of the range of
->
Replace motor
control (Pos. 1.35)
22
ROM error in the hand held programming unit:
Memory error in the hand held programming unit
->
Replace hand held programming unit (Pos. 1.31)
23
Invalid parameter:
Internal error in the hand held programming unit
->
Replace hand held programming unit (Pos. 1.31)
24
24V supply error motor control:
Error in the 24V supply in the motor control
->
Replace motor
control (Pos. 1.35)
->
Replace the power supply electronics (Pos. 1.41)
25
Bus error:
Bus system error
->
Replace spiral cable of the hand held programming unit
->
Replace hand held programming unit (Pos. 1.31)
->
Replace motor
control (Pos. 1.35)
26
24V supply hand held programming unit:
24V supply of the hand held programming unit is
defective
->
Replace hand held programming unit (Pos. 1.31)
27
5V supply hand held programming unit:
5V supply of the hand held programming unit is defective
->
Replace hand held programming unit (Pos. 1.31)
28
3.3V supply hand held programming unit:
3.3V supply of the hand held programming unit is
defective
->
Replace hand held programming unit (Pos. 1.31)
29
Calibration:
The calibration data in the motor control are wrong.
->
Perform a calibration
30
Calibration error:
->
Repeat the calibration
->
Replace knee electronics (Pos. 9.11)
->
Replace motor
control (Pos. 1.35)
31
Calibration timeout:
->
Replace motor
control (Pos. 1.35)
32
Motor enable error:
The motor could not be enabled
->
Replace motor
control (Pos. 1.35)
33
Motor disable error:
The motor could not be disabled
->
Replace motor
control (Pos. 1.35)
34
Motor stop error:
Motor stop command timeout error
->
Replace motor
control (Pos. 1.35)
35
Configuration error:
Invalid configuration of the hand held programming unit
->
Replace hand held programming unit (Pos. 1.31)
motion.
.
: