9) With the instrument in the VERTICAL position, biefly
press the button, the message Analyzing reference will
appear on the display;
10) Wait until the display displays the Insert sample
message;
Reading – Apparent Color
While Insert sample is on the display:
11) Open the measuring cover and remove the cuvette
from the compartment;
12) With the aid of the syringe, collect al least 15mL of the
sample to be analyzed;
13) With the sample in the syringe, fill in the glass cuvette
(12ml);
14) Close the cuvette with its cover;
15) Clean and dry puter wall of the cuvette using a cleaning
wipe or soft towel paper;
16) Insert the cuvette into the measuring compartment;
17) Measuring alignment: for better accuracy insert the
cuvette always in the same position;
18) Close the measuring cover on the glass cuvette to
avoid interference in the readings;
19) Briefly press the Power-Measuring button and wait.
The display will display the Message Analyzing
Sample, followed by the measurement result in PCU;
20) At the end of the measurement, discard the sample
analyzed and immediately wash the glass cuvette with
the distilled or deionized water.