Ametek Land SPOT Instrukcja instalacji - Strona 13

Przeglądaj online lub pobierz pdf Instrukcja instalacji dla Stojaki i stojaki Ametek Land SPOT. Ametek Land SPOT 16 stron. Water cooled mountings including sealed to process variant

Ametek Land SPOT Instrukcja instalacji
Water Cooled Mountings

3.1 Using an Analogue Cable

1)
For the Analogue Cable,
unscrew the Blanking Plug
from the Water Cooled
Mounting.
2)
Disassemble the M25 Cable
Gland. Note: The Hexagonal
Locknut is not required for this
installation.
3)
Slide the Analogue Cable
connector through the gland
outer shell and the seal
washer (Fig. 3-7).
4)
Fit the split bush between the washer and the main body of the gland.
5)
Insert the Analogue Cable into the connector on the rear of the
thermometer.
6)
Tighten the cable gland into the body of the thermometer.
7)
Close the hinged Back Cap and secure it in place with the fastener (Fig.
3-6).
The Thermometer is now installed in the Water Cooled Mounting.
Installation Guide
Fig. 3-7
SPOT
9