The chiller for the crystal is not running or does not have high enough flow rate. Check that
the chiller is on, the water level is high enough and that the filter at the back of the chiller is
not clogged. If the water is low, refill with a mix of clean water and 10 % Optishield 1.
1.1.1.3.
The key on the unit is OFF or the cable connection is bad, check that all cables are connected
properly.
1.1.1.4.
The humidity level in the diode is increasing, you need to change the desiccant on its side.
Turn off the diode unit, unplug the connectors at the back (not the fibre, be extra careful not
to pull on it), pull the unit in order to remove the lid. Unscrew the desiccant plug, put the new
one, close the lid, push the unit back in (pay attention not to bend the fibre when pushing)
and reconnect the cables. The desiccant plugs can be bought form Amplitude.
1.1.1.5.
The diode temperature will only turn on if the diode is not cooled properly, which should not
happen if the chiller is running unless one of the pipes inside the diode unit gets clogged
(which should also be prevented if the filters at the back of the chillers are cleaned regularly
and the water circuit flushed twice a year). To remove the blockage, you can reverse the water
flow several times for one minute at a time until the flow is back to normal. For the regen and
MP1 diodes the pipes to invert are located after the deionising unit. Be careful to put the
water flow in the correct direction.
Be careful to flush the water and clean the system afterwards according to the procedure of
the start-up manual.
1.1.1.6.
The diode unit is not receiving a stable trigger signal from the SelectPulse. Check that it is
working fine as well as the cable connections.
1.1.1.7.
The fibre is not plugged properly either on the diode or on the laser amplifier. This should not
happen as in normal operation the fibres are not to be removed from either side. Please
contact Amplitude.
1.1.1.8.
Hardware or software issue, please contact Amplitude.
1.2. Chillers
If an error occurs, it will be displayed on the LCD screen of the chiller.
1.2.1. Low water level
There is too little water in the tank. It must be refiled with the correct liquid (either
demineralised water or water + Optishield).
Note: if the water level drops faster than once every few months, the water connections
should be checked as there might be a small leak.
3
Magma 200
Quick troubleshooting guide
Version 1.00
Device
Human
Fiber humidity
Diode temperature
Trigger
Fiber connection
Internal
Amplitude Laser Group
11 avenue de Canteranne, Cité de la Photonique - Bâtiment MEROP A 33600 Pessac - France
Headquarters
/
Capital social: 317 100 €
/
/
Siret 434 637 005 - APE 6420Z
Tel: +33 (0)5 56 46 40 60
/
Fax: +33 (0)5 56 46 06 94
Headquarters
/
11 avenue de Canteranne, Cité de la Photonique - Bâtiment MEROPA 33600 Pessac - France
Capital social : 320 925 €
Amplitude Laser Group
/
Tel: +33 (0)5 56 46 40 60
/
N°TVA Intracom: FR 70 434 637 005
/
/ [email protected] / www.amplitude-laser.com
/
SIRET: 434 637 005 - APE 6420Z
/
N°TVA Intracom: FR 70 434 637 005
Fax: +33 (0)5 56 46 06 94
/
/ www.amplitude-laser.com