Warnings
a. Allow at least 20 minutes after eating or drinking before testing. Any type of
alcohol remaining in the mouth of saliva may interfere with proper testing
and requires about 20 minutes to disperse. This includes alcohol-based
mouthwash or similar products.
b. Avoid testing in high wind conditions or in poorly ventilated or enclosed spaces.
Poor ventilation may also lengthen the time required for warm up between tests.
c. Do not blow smoke, saliva or other contaminants into the mouthpiece, as the
sensor may become damaged. Wait at least 20 minutes after smoking before
conducting a test.
d. Do not use the breathalyser as a tool to drink and drive.
e. Do not blow fluid into the instrument as it will damage the sensor and void your
warranty.
f. Do not store in an extreme environment where high temperature or high
humidity exists. Keep out of the reach of children.
g. To maintain accuracy, we recommend professional recalibration every 6 months.
h. Blood alcohol concentration can continue to rise for up to 2 hours after the
cessation of drinking. If a caution result or a result is close to the designated
BAC, it is recommended to wait and retest after 20 minutes.
i. It can take 10 hours or more for the alcohol level to return to a zero after a high
blood alcohol level has been reached. A further test should be carried out later
in the day of the following morning.
Recalibration & Service
The AlcoSense
Nexus uses a fuel cell sensor to measure breath alcohol levels.
®
Depending upon the amount of alcohol registered by the sensor after a period of
time, it will require recalibration.
Andatech recommends the AlcoSense
or after 500 uses, whichever comes first. To maintain it to Australian Standards,
calibration every 6 months is required.
Nexus to be calibrated every 6 months
®