Alert-it Care Systems P163BAB Podręcznik - Strona 6
Przeglądaj online lub pobierz pdf Podręcznik dla System bezpieczeństwa Alert-it Care Systems P163BAB. Alert-it Care Systems P163BAB 10 stron. General alarm monitor
Battery Replacement
Expected Battery Life: Operating Mode
No failsafe heartbeat, no alarms
Fail safe heartbeat, and alarms
Maximum alarm number (based on 2 min response)
The battery needs replacing when the alarm led flashes intermittently every 8
seconds. The P163 will continue to function for some hours, allowing time for this
replacement. If the battery is exhausted then the P163 will click continuously
without the light flashing.
Battery Type: 2 of AA Alkaline Batteries or equivalent
Cleaning:
The following is a general guide line based on the components listed on page 3. Where a
different sensor has been supplied, then please refer to the cleaning instructions supplied
with that sensor.
Technique A
Wetting with strong disinfectant. This can include immersion provided plugs and any
obvious breathing holes are avoided.
Technique B
Wiping with cotton wool pads moistened (compressed until dripping stops) with a
mild detergent (0.5% washing up liquid) solution.
Technique C
Wiping with disposable 70% isopropyl alcohol wipes. Ensure that any plugs are
completely dry before reinserting into the sensor input socket on the monitor.
Technique D
Full immersion in detergent , water and optional disinfectant. See component
washing instructions for details
UH1142B P163 Wired General Interface v2 page 6 of 8
Maintenance
Battery capacity
1 year
6 months
2000 alarms