Apogee SQ Series Instrukcja obsługi - Strona 6

Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Akcesoria Apogee SQ Series. Apogee SQ Series 19 stron. Quantum sensor

SPECIFICATIONS

Power Supply
Sensitivity
Calibration Factor
(Reciprocal of Sensitivity)
Calibration Uncertainty
Calibrated Output Range
Measurement
Repeatability
Long-term Drift
(Non-stability)
Non-linearity
Response Time
Field of View
Spectral Range
Directional (Cosine)
Response
Temperature Response
Operating Environment
Dimensions
Mass
Cable
Calibration Traceability
Apogee SQ series quantum sensors are calibrated through side-by-side comparison to the mean of transfer
standard quantum sensors under a reference lamp. The reference quantum sensors are recalibrated with a 200 W
quartz halogen lamp traceable to the National Institute of Standards and Technology (NIST).
± 5 % (see Calibration Traceability below)
Less than 1 % (up to 4000 µmol m
370 to 650 nm (wavelengths where response is greater than 50 % of maximum; see Spectral
± 5 % at 75° zenith angle (see Cosine Response below)
10 to 60 C; 0 to 100 % relative humidity; can be submerged in water up to depths of 30 m
5 m of two conductor, shielded, twisted-pair wire; TPR jacket (high water resistance, high UV
stability, flexibility in cold conditions); pigtail lead wires; stainless steel (316), M8 connector
SQ-100X-SS
Self-powered
-2
-1
0.1 mV per µmol m
s
-2
-1
10 µmol m
s
per mV
0 to 400 mV
Less than 0.5 %
Less than 2 % per year
Less than 1 ms
180°
Response below)
- 0.04 % per C
24 mm diameter, 33 mm height
90 g (with 5m of lead wire)
located 25 cm from sensor head
-2
-1
s
)