Always remove the rechargeable batteries
if you do not intend to use the monitor for a
week or more.
The batteries can NOT be charged while inserted
in the blood pressure monitor. Recharge the
batteries in an external charger and observe
the information regarding charging, handling
and service life.

6. Using a mains adapter

You can operate this monitor using the Microlife
mains adapter (DC 6 V, 600 mA).
Only use the aponorm
available as an original accessory appropriate
for your supply voltage. You can purchase the
mains adapter from your local pharmacy.
Ensure that neither the mains adapter nor the
cable are damaged.
1.
Plug the adapter cable into the mains adapter
socket
of the blood pressure monitor.
7
2.
Plug the adapter plug into the wall socket.
When the mains adapter is connected, no battery
current is consumed. We recommend to have
batteries in the monitor even with mains operation,
just as a back-up. They ensure that in the event of
a power failure date and time are retained and do
not have to be entered manually again.
12
mains adapter
®
7. Error messages
When an error occurs during a measurement, the
measurement is interrupted and an error message
displayed, e.g. «Err 3».
If you do not receive an error message, but if
the results seem unusual to you, please check
whether you observed all points of the check
list on page 8 and in the enclosed flyer "8
golden rules of taking the blood pressure".
For more detailed help in troubleshooting the
monitor, please refer to the download section
of the www.aponorm.de product website.
Error
Designa-
tion
«ERR 1»
Signal too
-E
weak
17
Potential cause and
remedy
The pulse signals on
the cuff are too weak.
Re-position the cuff and
repeat the measure-
ment.*