ADE HWZ Series Instrukcja obsługi - Strona 11

Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Wagi ADE HWZ Series. ADE HWZ Series 20 stron. Electronic counting scale

ADE HWZ Series Instrukcja obsługi
The functions F1 CAL and F8 GrV are only reserved for specialist personnel authorised by the
manufacturer.
Accidental or intentional changes to these parameters by unauthorised persons can sig-
nificantly affect the scale functionality
"Weight"
display
F2 DI
F3 CnT
F4 AU
F5 AZN
F6 PLn
F7 SPD
F9 Off
F10 Bep
F11 RSt
HWZ-190214-Rev006-UM-en
Description
Sets the scale step. Press
key to scroll through the options (1, 2, 5, 10, 20 or 50). Press
select an option
_______ Displays the number of A/D conversions. Press
the menu.
Press
to view the currently set value. Press the
through the options. Use the
Au on
For automatic totalling and printing when the scale is stable,
Au oFF
For manual totalling. To total data, press
P Cont
For continuous printing without totalling.
The optional RS232 data interface and a connected printer are required for
printing.
Press
to view the auto-zero range options. Press
rently set value. Press the
value can be set to 0.5d, 1d, 2d or 4d. Normally it is set to 1d, but it can be
increased in order to zero the scale if there is a small amount of material
spilled on the weighing platform during operation.
Sets a new code. Press
then press the
key. "____" is displayed. Enter the code again and con-
firm with
. "_____" is displayed: The new code was accepted. Keep the
new code in a safe place for later use!
Adjusts the speed of the A/D converter. The settings are 7, 5, 15, 30 and 60.
The slowest settings are 7, 5 and the fastest one is 60.
Automatic switch-off setting. The scale can be set to switch off after 3/5/15 or
30 minutes or the switch-off can be deactivated. Press the
change the setting.
Setting the acoustic alarm. Options:
OK:
Acoustic alarm sounds if the entered value is within the limit values.
Low:
Acoustic alarm sounds if the entered value is below the minimum
value.
NG:
Acoustic alarm sounds if the entered value is outside the limit values.
High:
Acoustic alarm sounds if the entered value is above the maximum
value.
Resets the scale to its original state.
to view the currently set value. Press the
key to select:
key to navigate through the options. The
. "____" is displayed. Enter the new code and
to
to return to
key to navigate
to view the cur-
key to
11