Great Life Healthcare UTS Podręcznik użytkownika - Strona 13

Przeglądaj online lub pobierz pdf Podręcznik użytkownika dla Sprzęt medyczny Great Life Healthcare UTS. Great Life Healthcare UTS 18 stron.

Great Life Healthcare UTS Podręcznik użytkownika
The Universal Therapy System (UTS) is intended for use in the electromagnetic environment (for home and
professional healthcare) specified below.
The customer or the user of the Universal Therapy System (UTS) should assure that it is used in such and
environment.
Immunity test
IEC 60601 test level
Conducted RF
3 Vrms:
IEC 61000-4-6
0,15 MHz – 80 MHz
6 Vrms:
in ISM and amateur
radio bands between
0,15 MHz and 80 MHz
80 % AM at 1 kHz
Radiated RF
10 V/m
IEC 61000-4-3
80 MHz – 2,7 GHz
80 % AM at 1 kHz
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
Manufacturer's declaration-electromagnetic immunity
Compliance level
3 Vrms:
0,15 MHz – 80 MHz
6 Vrms:
in ISM and amateur
radio bands between
0,15 MHz and 80 MHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Electromagnetic environment-guidance
(for home and professional healthcare
Portable and mobile RF communications
equipment should be used no closer to
any part of the Universal Therapy System
(UTS) including cables, than the
recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance:
d = 1,2 √P
d = 1,2 √P 80MHz to 800 MHz
d = 2,3 √P 800MHz to 2,7 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the recommended
separation distance in meters (m).
Interference may occur in the vicinity of
equipment marked with the following
symbol:
13
environment)