General requirements for basic safety and essential performance - Collateral standard:
Electromagnetic compatibility - Requirements and Tests),IEC 60601-1-11 (Medical
electrical equipment – Part 1-11: General requirements for basic safety and essential
performance – Collateral Standard: Requirements for medical electrical equipment and
medical electrical systems used in the home healthcare environment),the ASTM
(American Society for Testing and Materials) E 1965 - 98 , ISO 80601-2-56 ( Medical
Electrical Equipment -- Part 2-56: Particular Requirements For The Basic Safety And
Essential Performance Of clinical thermometers for body temperature
measurement).Please note that portable and mobile HF communication systems may
interfere with this unit.
This infrared thermometer meets requirements established in ASTM Standard (E1965-98)
except of clause 5.2.2. It's display range is 34.0℃-43.0℃(93.2℉-109.4℉).the Full
responsibility for the conformance of this product to the standard is assumed by
manufacturer.
ELECTROMAGNETIC COMPATIBILITY INFORMATION
Phenomenon
Conducted
radiated RF emissions
Harmonic distortion
Voltage
fluctuations
and flicker
Phenomenon
Electrostatic
Discharge
Radiated RF EM
field
Proximity fields from
RF wireless
communications
equipment
Rated power
frequency magnetic
fields
Table 1 – Emission
Compliance
and
CISPR 11
Group 1, Class B
IEC 61000-3-2
NA
IEC 61000-3-3
NA
Table 2 – Enclosure Port
Basic EMC
standard
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-3
IEC 61000-4-8
Electromagnetic environment
The device is intended to be used in
home healthcare environment
The device is powered by battery
The device is powered by battery
Immunity test levels
Home Healthcare Environment
±8 kV contact
±2kV, ±4kV, ±8kV, ±15kV air
10V/m
80MHz-2.7GHz
80% AM at 1kHz
Refer to table 3
30A/m
50Hz or 60Hz
11