OTDR Mode: Event Table Page Features
ICON
EVENT TYPE DESCRIPTION
Macrobend
Splitter
Group Start
Event
Group
Middle Event
Group End
Event
OTDR Mode: Summary Page Features
Summary page displays:
trace graph, fiber under test [Length] in user
selected units, [Loss] and [ORL] in (dB).
For multiple-wavelength tests, press to toggle
the wavelength and display [Length], [Loss]
and [ORL] test results for that wavelength.
A non-reflective loss event with significantly higher loss at longer wavelengths (1550,
1625, 1650 nm) than seen at shorter wavelengths (1310, 1490 nm). Typically caused
by a sharp bend in the fiber.
PON splitters are detected only when Event thresholds are set to PON Default or PON
User and a high loss event (> 6 dB) is detected which is determined not to be the end
of the fiber.
First event in multiple event group. Reports loss of entire group. Reports
reflectance of first event in group if reflectance of next event can be independently
measured. Reports max reflectance of grouped events if next reflectance cannot be
independently measured.
Grouped events may contain zero, one, or more middle events. Loss of middle events
is included in group loss reported in the group start event. Reports reflectance if it
can be independently measured.
Last event in a multiple event group. Loss of last event is included in group loss
reported in group start event. Reports reflectance if it can be independently
measured.
11