ASSEMBLY
|
Fig. 1
3
²
@
³
Mounting on the Microscope
1. Rotate the coarse adjustment knob @ to fully raise the stage.
# Be careful not to raise the stage too much to avoid acci-
dental contact with the objective lens and specimen.
2. Lower the condenser holder with the condenser height ad-
justment knob ² until it stops at the lower limit.
3. Loosen the condenser clamping knob ³.
4. Insert the phase contrast turret condenser into the dovetail
guideway of the condenser holder until the positioning pin of
the condenser is engaged in the positioning groove of the
mounting dovetail guideway.
5. Tighten the clamping knob of the condenser holder.
6. Raise the condenser holder until it stops at the upper limit.