Abbott i1000SR Skrócona instrukcja rozwiązywania problemów - Strona 2

Przeglądaj online lub pobierz pdf Skrócona instrukcja rozwiązywania problemów dla Sprzęt laboratoryjny Abbott i1000SR. Abbott i1000SR 2 stron.

Abbott i1000SR Skrócona instrukcja rozwiązywania problemów
i1000SR System Quick Troubleshooting Guide
Customer Support: 1-877- 4ABBOTT (1-877- 422-2688
Level Sense Errors: 3000
• Ensure sample cup or tube is present and adequate
sample is in the cup or tube.
Reagent probe areas only:
• Reagent bottle is empty: Replace empty reagents.
• WZ probe maintenance water bottle is missing, over-
filled or filled with distilled (DI) water instead of tap
water or saline or bleach: Repeat the maintenance
procedure, ensuring the WZ probe maintenance water
bottle is in position and contains the correct volume
and solution.
Only fill maintenance
bottle 25 to 30ml full.
Errors may occur if
filled with more than
30ml of solution.
Sample Pipettor:
Tighten the probe clamp finger tight.
PN 205854-101
Error Code 5209 No Reaction Vessel (RV)
Detected in Process path after RV Load
Attempt
• If RV loader chute is not in position to load RVs,
©2009, Abbott Laboratories, Abbott Park, IL, 60064
Refer to the Operations Manual for further instructions
)
Instrument SN: ______________
reseat the RV loader chute into the wash zone
manifold.
Chute is in the UP position (incorrect).
Chute is in the DOWN position (correct).
®
Page 2 of 2