AmeriWater 00HC-0093 Manual de operação e manutenção - Página 7
Procurar online ou descarregar pdf Manual de operação e manutenção para Aquecedor de água AmeriWater 00HC-0093. AmeriWater 00HC-0093 10 páginas. Healthcare storage tanks
2.0
INSTALLATION AND STARTUP
1.
Locate the storage tank on a firm, level foundation. For seismic
requirements, drill (4) 5/8" diameter holes into the concrete through
the mounting holes in the storage tank feet a minimum of 4" deep.
Install (4) 5/8" diameter, HILTI KB-TZ Expansion anchors through
the 4 holes on the pads to anchor into the ground. A minimum of 4
threads for each anchor must be below the concrete prior to
application of 60 ft-lbs of torque.
2.
Connect the discharge from the pump to the distribution piping.
3.
Run the floatswitch quick disconnect to the RO and lock in place.
4.
Prime the pump by filling the tank with water and opening the valve into to
the pump's inlet. Loosen the priming plug on the pump to allow the air to
escape. Once water starts to escape, re-tighten the priming plug
(Only applicable to pump 80-0002).
5.
Wire or plug the storage tank power cord into a dedicated UL listed
NEMA 4X rated disconnect. See table below for models and pump
voltages.
Part #
Electrical Voltages
80-0002
115VAC, 1PH, 9.2AMPS
80-0148
220VAC, 1PH, 7.5AMPS
80-0203
208VAC, 3PH, 8.3AMPS
6.
If the pressure delivered by the pump is to be changed, a technician will
need to have access to a smartphone with the Grundfos GO application
along with a Grundfos communication dongle (AMW PN 97-0003). Refer
to IFU 98-9057 to setup parameters to change pressure delivered by the
pump.
7.
Loop should be flushed prior to operating the distribution pump on the
storage tank.
Name
DISTRIBUTION PUMP-55,100
& 185 GALLON STD FLOW,
MODELS 00HC-0090, 00HC-
0090-PRV, 00HC-0092 AND
00HC-0094
DISTRIBUTION PUMP, HIGH
FLOW 30GPM,
MODELS 00HC-0091, 00HC-
0093, 00HC-0095
DISTRIBUTION PUMP, HIGH
FLOW 70GPM,
MODEL 00HC-0097
5