VII. ELECTRICAL SPECIFICATIONS
Ground Resistance (Auto range) :
Range
0.025 - 0.250 
0.250 - 1.000 
1.001 - 9.999 
10.00 - 50.00 
10.01 - 99.99 
100.0 - 200.0 
200.1 - 400.0 
400.0 - 600.0 
600.1 - 1500 
Loop resistance noninductive, external field < 50 A/m, external electrical
1
field < 1 V/m, conductor centered.
Resistance Measurement Frequency: 3.333KHz
2
High and Low Alarm
High Alarm
Lo Alarm
Ground/Leakage Current
(Auto range, 50/60 Hz, True RMS, Crest Factor < 3.0)
Range
0.200 - 1.000 mA
1.00 - 10.00 mA
10.0 - 100.0 mA
100 - 1000 mA
(Auto range, 50/60 Hz, True RMS, Crest Factor < 3.0)
Range
0.200 – 4.000 A
4.00 – 30.00 A
Accuracy of Resistance Calibration Plate: ±0.5%
Data Logging Capacity
Data Logging Interval: 1 to 255 seconds
GlobalTestSupply
www.
Resolution
Accuracy of Reading
0.002 
± 1.5% ± 0.05 
0.02 
± 1.5% ± 0.05 
0.02 
± 1.5% ± 0.1 
0.04 
± 2 .0% ± 0.3 
0.04 
± 2 .0% ± 0.5 
0.4 
± 3.0% ± 1.0 
2 
5 
20 
Range
0 - 1510
0 - 1510
Resolution
Accuracy of Reading
0.001 mA
± 2.0% ± 0.05 mA
0.01 mA
± 2.0% ± 0.03 mA
0.1 mA
± 2.0% ± 0.3 mA
1 mA
Resolution
Accuracy of Reading
0.001 A
± 2.0% ± 0.03 A
0.01 A
± 3.0% ± 0.03 A
116 records
14
1
± 5.0% ± 5 
± 10% ± 10 
± 20%
Resolution
1
1
± 2.0% ± 3 mA
.com