4
Locate a vial of the Secondary Verification Standard you chose from
the bag of standards or in a separate box you've purchased.
5
Twist off the vial cap and pour the contents into a new sample cup.
6
Take a reading of the second standard by placing the sample cup
into the chamber, closing the drawer, and moving the knob to the
left (Read). Take two readings.
Dew Point Mode: If you are using the dew point block that comes standard with the
Pre and the value of the second verification is within +/-0.01, it is safe to start testing
your product(s).
Capacitance Mode: If you are using a volatiles-capacitance block in your Pre and
the value of the second verification is within +/-0.015, it is safe to start testing your
product(s).
Standard not measuring within spec: If the instrument doesn't measure within
specifications, clean the chamber and retry. Contact AquaLab Customer Support at
[email protected] or 509-332-5601 if you need any assistance.