Olympus DP30BW Manual de instruções - Página 5
Procurar online ou descarregar pdf Manual de instruções para Câmara digital Olympus DP30BW. Olympus DP30BW 31 páginas. Microscope digital camera
1
Conformity of the System
Restrictions
1. The recommended C-mount camera adapter is the combination U-TV1X-2 + U-CMAD3 or the U-TV0.5XC-2. (The
U-TV0.5XC should not be used because it deteriorates the image flatness.)
A TV adapter with a magnification below 0.5X cannot be used because it obscures part of the image.
2. When combination SZX + SZX-RFA (RFL) + SZX-BS + SZX-PHA is used, the DP30BW can be used only when it is
mounted on the port on the opposite side to the lamp housing.
When combination SZX + SZX-RFA (RFL) + SZX-TR30 is used, the DP30BW can be used only when a large stand or
illumination stand is used.
3. When a double-port type intermediate attachment (U-DP, U-TRU, etc.) is used, only the port coming on the optical axis can
be used.
4. When the U-TV0.5XC-2 is used together with a low-power objective (1.25X or 2X), the peripheral part of image may be
obscured unless the U-ULC-2 ultralow-magnification condenser is used.
5. When the DP30BW is connected to the rear port of the U-DPT or U-MPH, the peripheral part of the image may be
deteriorated due to the optical performance of the U-DPT or U-MPH.
6. When the U-CFU is used, it is required to set the exposure to a longer period than 1/30 sec. using the manual exposure
mode and adjust the brightness by engaging or disengaging ND filters.
7. Intermediate attachments:
When the U-TV0.5XC-2 is used, using two or more intermediate attachments* may obscure the peripheral part of the field
of view.
* Two intermediate attachments: BX51 + U-URA + U-CA or equivalent.
8. A commercially available C-mount lens can be used provided that it is designed for a 2/3 in. or larger CCD and the lens
projection from the C-mount area is less than 4.5 mm. However, Olympus does not guarantee the performance in this
case.
If a lens that does not employ a telecentric optics is used, problems due to optical adaptability such as shading may
occur.
Operating Environment
Temperature: 10 to 35°C (50 to 95°F). Humidity: 20 to 85% (without condensation).
DP30BW
2