CKX31/CKX41
Parfocality adjustment amount (CKX31: reference value)
Reading value
(before adjustment)
AB129900 AB130000 AB130200 AE092100 AE092200 AE092300
2.0
1.9
1.8
1.7
1.6
1.5
1.4
1.3
1.2
1.1
1.0
0.9
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0.0
-0.1
-0.2
-0.3
-0.4
-0.5
-0.6
-0.7
-0.8
-0.9
-1.0
-1.1
-1.2
-1.3
-1.4
-1.5
-1.6
-1.7
-1.8
-1.9
-2.0
Reading value (before adjustment): Reading value of parfocality difference on the helicoid scale of standard eyepiece
Reading value (after adjustment): Theoretical reading value of parfocality difference after adjustment
Ex.) Parfocality differnce is 1.9 mm: when the total thickness (1.70mm) of washers, (0.7mm: 1pc.) and ( 1mm: 1 pc.),
* Numeric character (1 or 2) in the column of washer selection shows the number of washers.
D. REPAIR PROCEDURE
Washer selection
0.05
0.1
0.3
1
1
1
1
1
1
1
1
1
1
1
1
2
1
2
2
1
1
1
1
1
1
2
2
2
1
1
1
1
1
2
1
1
1
1
1
1
1
1
1
1
2
are inserted , parfocality difference becomes 0.06mm reading value.
< standard : 1.0 >
0.7
0.5
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
D-7
Unit: mm
Total
Reading value
washer
(after adjustment)
thickness
1.0
1.75
0.03
1
1
1.70
0.06
1
-0.04
1.70
1
1.65
0.00
1
0.03
1.60
1
1.55
0.06
1
-0.04
1.55
1
1.50
-0.01
1
1.45
0.02
1
1.40
0.05
1
1.40
-0.05
1
1.35
-0.02
0.01
1
1.30
1
0.04
1.25
1.25
-0.06
1
1
1.20
-0.02
1
1.15
0.01
1.10
0.04
1
1.10
-0.06
1
1
1.05
-0.03
1
1.00
0.00
0.95
0.03
0.90
0.06
0.90
-0.04
0.85
-0.01
0.80
0.02
0.75
0.06
0.75
-0.04
0.70
-0.01
0.70
-0.11
0.60
0.05
0.60
-0.05
0.55
-0.02
0.50
0.01
0.50
-0.09
0.45
-0.06
0.40
-0.03
0.35
0.00
0.30
0.04
0.30
-0.06
0.20
0.10