Olympus SZX16 Manual de instruções - Página 17
Procurar online ou descarregar pdf Manual de instruções para Microscópio Olympus SZX16. Olympus SZX16 28 páginas. Research high-class stereo microscope
Também para Olympus SZX16: Visão geral (24 páginas), Manual de instruções (36 páginas), Manual de instruções (40 páginas)
6
Fig. 15
A
B
C
D
SZX16 REFLECTED FLUORESCENCE SYSTEM
4. Rotate the collector lens focusing knob 6 to project the arc image on
the stage plate. (A)
7
If the arc image is not projected, adjust the burner centering knobs 7.
5. Rotate the burner centering knobs 7 to move the arc images on the 45°
position on the top right (or bottom left) of the illumination field. (B)
6. Insert the Allen screwdriver into the mirror focusing screw 8 (Fig. 16) and
turn the screw to focus on the mirror arc image. (C)
7. Rotate the burner centering knobs 7 to overlap the arc image with the
mirror arc image. (D)
}After starting observation, rotate the collector lens focusing knob 6 to
make the observation field uniform.
}Hereafter, the mercury burner does not need centering until the next time
it is replaced.
}When the objective is removed, more accurate centering of the mercury
burner is possible. At this time, set the zoom knob magnification at around
3.2X.
14