with preeclampsia.
This product might not meet its performance specifications if stored or used outside the specified
temperature and humidity ranges.
Please do not share the cuff with any infectious person to avoid cross-infection.
11
The monitor, cable, battery and cuff must be disposed of according to local regulations at the end of their usage.
7. Prolonged over-inflation of the bladder may cause bruises of your arm.

10. CARE AND MAINTENANCE

1. Do not drop the monitor or subject it to strong impacts.
2. Avoid high temperature and direct exposure to the sun. Do not immerse the monitor in water.
3. If this monitor is stored near freezing temperatures, allow it to acclimate to room temperature before use.
4. Do not attempt to disassemble this monitor.
5. If you do not use the monitor for a long time, please remove the batteries.
6. It is recommended that the performance of this monitor be checked every 2 years.
7. Please clean monitor with a soft dry cloth. Do not use any abrasive or volatile cleaners.
8. The supplier will make available on request the circuit diagrams, component part lists, descriptions,
calibration instructions, or other information which will assist the user's appropriately qualified technical
personnel to repair those parts of equipment which are designated by the manufacturer as repairable.
9. It is recommended that if the cuff is used in a hospital or a clinic, it be disinfected twice a week. Wipe the
inner side (the side that contacts skin) of the cuff with a soft cloth lightly moistened with Ethyl alcohol
(75-90%). Then air dry the cuff.
10. The monitor can maintain the safety and performance characteristics for a minimum of 10,000
measurements or three years, and the cuff integrity is maintained after 1,000 open -close cycles of the
closure.