Atlantic Ultraviolet Sanidyne 40-0674A Руководство пользователя - Страница 9

Просмотреть онлайн или скачать pdf Руководство пользователя для Аксессуары Atlantic Ultraviolet Sanidyne 40-0674A. Atlantic Ultraviolet Sanidyne 40-0674A 20 страниц. Uv portable area sanitizer with remote

Atlantic Ultraviolet Sanidyne 40-0674A Руководство пользователя
6. Start Button: Starts treatment cycle time.
Once selected:
a) Remaining Time changes to Countdown Time and begins counting down from 1:29 minutes to exit area
b) The words "LEAVE AREA" appear on the lower status bar.
c) Alarm on unit begins to sound
d) The word "PREP" flashes on the Keypad during the Countdown Time, when Treatment Time begins the word "RUN"
flashes and when the treatment time is completed the word "DONE" will appear.
7. Return Button: Brings you back to previous screen
8. Pause / Continue Button: Used to pause the treatment time, or to continue the treatment time. This button toggles from the
Pause Icon, which is displayed as two vertical bars to the Continue Icon which is a looping arrow. If pause is selected below
30 seconds during the Countdown, the Countdown time will reset to 30 seconds. If pause is selected during the cycle time the
device will provide an additional 1:30 seconds of countdown time to leave the area.
9. Battery Condition: Displays the charge level of the battery in the hand-held remote. Press on battery symbol to see
percent of change.
10. Settings Button: Contains choices for:
a) Settings: Reset Operating Time or Clear Session Data
b) Set Date/Time: requires a PIN to reset (See "Setting Date and Time" on Page 12)
c) Sessions: Log of saved sessions
11. Remaining Time: Displays remaining time of treatment and toggles to (Countdown time shown on Treatment in Hours,
Minutes, Seconds) when selected.
12. Total Operation Time: Displays accumulative hours of all treatment times.
13. Stop or End Button: Stop treatment time.
14. Status Bar: Outputs all the conditions of the device
a) Ready
b) Intruder Alert
c) Session in Progress
d) Session Paused
e) Session Resumed
f) Session Stopped
g) Leave Area
h) Connected to Sanidyne
15. Settings Button: Same as (10) above.
16. Home Button: Exit Setting
l
i
amP
nstallation
Always disconnect power to the fixture before performing any service or maintenance.
Germicidal lamps are easily damaged. Exercise care when handling lamps.
Always Disconnect Power to the
Prime Remote before
Sanidyne
®
installing or removing lamps.
®
Step 1 - Insert lamp end without pins
through BOTTOM of lamp guard while
holding each lamp end with one hand.
Step 2 - Carefully align pins on lamp
base to socket holes on lampholder
as shown. (Lamp Installation
Continued on Page 10.)
9