7-4.Set comparator LO (Initial value: 0000)
〇 TRACK / PEACK Mode
・You can change the setting value on the LO side of comparator.
*Refer to "7-3. Set comparator HI" for the function description, setting method, and operation method.
〇 PP Mode
・You can change the setting value on the - side (lower section) of comparator.
For example, if you set the point of + side (HI) to 500N and the point of - side (LO) to -100N, and
gradually apply force to the - side from 0N, the display of the lower section will be changed, and
when the force exceeds -100N, a short intermittent sound will be heard.
* The setting value should be entered as the value of - side instead of a absolute value.
* A short intermittent sound will be heard if the HI setting value (+ side) is exceeded, and a long one if the LO
setting value (- side) is exceeded. After that, even if the HI peak value is exceeded, there is still a long
intermittent sound.
* A long intermittent sound will be heard if the LO setting value (- side) is exceeded, and a short one if the HI
setting value (+ side) is exceeded. After that, even if the LO peak value is exceeded, there is still a short
intermittent sound.
* The position of the decimal point varies with each unit and each model.
* For models that increase in digits, the maximum setting value is 9999.
 For example: When the unit is N in AD-FG10, the settable maximum value is -99.99(N).
* The setting value returns 0 when the mode is changed.
* Short press the "FUN" switch to confirm the setting. Next is the setting for overload.
7-5.Set overload OV (Initial value: 120)
・It is a function to warn the users to avoid accidental damage to the load sensor.
For example, when this gauge is set on a load stand to perform a test, it can display a warning at 80% of full
scale and output a signal to stop the test so that the sensor is not damaged by applying excessive force
accidentally.
* The setting can be set by 10% up to 120%.
* Short press the "FUN" switch to confirm the setting. Next is the setting for screen update rate.
7-6.Set screen update rate UD (Initial value: 20)
・You can set the screen update rate (switching speed).
 By decreasing the value, you can check the changing values in more real time.
However, if the update rate is too fast, the values will change quickly and it may be difficult to see.
* Update rate is in seconds. If the display value is 20, the display will be updated every 20 seconds.
* There are 4 types of settings: 20, 10, 5, and 2 seconds.
* Short press the "FUN" switch to confirm the setting. Next is the setting for flipping screen.
"POW" Switch
"POW" Switch
Increase by 10 every
"POW" Switch
time the "POW" switch
is pressed.
"POW" Switch
"POW" Switch
"POW" Switch
"POW" Switch
"POW" Switch